HC997 – MOD. B
Digital multifunction solenoid metering pump
4.4.9
Proportional functioning by Rx signal
ADSP9000030
In this functioning, the parameters to set are the following:
Setpoint Rx
Indicates the Rx value to maintain constant.
Setting from 0 to 1000 at step of 1.
Prop. L1 Rx
Indicates the Rx value that together with the Setpoint value make the
proportional dosing band.
Setting from 0 to 1000 at step of 1.
% SPM Rx=Set
Indicates the dosing frequency of the pump at the Setpoint value.
Setting from 0 to 100%.
% SPM Rx=L1
Indicates the dosing frequency of the pump at the L1 value.
Settings from the set value on %SPM Rx=Set to 100%.
Cal. Rx
Perform the gain calibration of the Rx sensor, put the sensor inside a known
buffer solution (i.e. mV 465).
Press the
Qual. Rx Pr.
It's the percentage of the sensor's quality.
Restore Par.
Restoring the default parameters of the Rx functioning.
Press the
Press the
key to go back at the previous menu.
4.4.9.1 Diagram of the proportional functioning by Rx signal
PROPORTIONAL FUNCTIONING BY Rx SIGNAL
keys to adjust the measured value.
keys to set the desired values.