Step 2
While the
symbol blinks, press and hold the
< button for 3 seconds until 'ctl' appears on the
display. If the talking function is on, the meter
will say "Control solution test. Please apply
control solution onto the test strip. "
Step 3
Shake the ReliOn™ Premier Control
Solution bottle well before each
test. Remove the cap and squeeze
the bottle to discard the first 1 or
2 drops.
Then wipe the tip with a clean tissue
or cloth. While the
apply the control solution to the narrow edge of the test strip
until the meter beeps. Make sure the confirmation window of
test strip fills completely.
NOTE
• The meter may switch off if the control solution is not applied
within 2 minutes of the
the screen.
• If the meter turns off, remove the strip, reinsert and start
from step 2.
symbol blinks,
symbol appearing on
21