Operating Manual
Introduction
It is highly recommended to read this operating manual before using the
HM6050-2. After unpacking the equipment check for any mechanical
damage or loose parts inside. Should there be a transportation damage,
inform the supplier immediately. Do not put the equipment into operation.
Symbols as marked on equipment
Operation according to rules
The Line Impedance Stabilization Network HM6050-2 has to be operated
according to the regulations as defined in VDE 0876 Part1 ("Measurement
of radiated emissions"). It conforms to the regulations as defined in
CISPR Publ.16 or. EN55011.
Safety
This instrument has been designed and tested in accordance with VDE
0411 Part 1, Safety requirements for electrical equipment for
measurement, control, and laboratory use. The CENELEC regulations
EN 61010-1, international standard IEC 1010-1 res. correspond to this
standard. All case and chassis parts are connected to the safety earth
connector of the power line plug, corresponding to the Safety Class 1
regulations.
To put the HM6050-2 to operation it has to be connected to an AC
power outlet (^ with protective ground) according to VDE-0100. Due
to the relative high leakage current (ca. 800 mA) the equipment
cannot be operated in combination with a leakage current circuit
breaker. Thus, "protective ground" is equivalent to "earth".
Any interruption of the protective ground conductor inside or outside
the instrument is prohibited.
22
ATTENTON – refer to manual
Protective ground (earth) terminal
L1/N – identification of power lines
(Lit, if the power cable connector is plugged
in correctly (phase))
Subject to change without notice