Abbott Patrol Manual Del Usuario página 7

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

CONTROL DIAL SETTINGS (continued)
DIAL:
HOLD
VOL FED
CLEAR
VOL
FUNCTION:
USING THE HOLD FUNCTION
HOLD stops pumping action. Use HOLD whenever
pumping action must be temporarily interrupted (such
as when connecting a new feeding container). If pump
is left on HOLD for more than 5 minutes, an alarm
will sound and the message SEL run (SELECT run)
will be displayed.
NOTE: The HOLD timer may be programmed for a
duration of 1–90 minutes (in 1-minute intervals) by
depressing the keypad arrows either up or down when
the dial is in the HOLD position.
USING THE HOLD FUNCTION TO SILENCE AN ALARM
HOLD also is used to silence an alarm. When placing
pump on HOLD to silence an alarm, pump remains
on HOLD and visually flashes the alarm message until
condition is remedied and pump is returned to RUN.
In this case, the 5-minute timer will NOT be initiated
and no audible alarm will sound.
Stops pumping action. The display will show the
volume delivered. Pump accumulates volume delivered
up to 9999 mL unless the volume is reset to zero by
turning to CLEAR VOL (or the pump is unplugged
while in the OFF/CHARGE mode). Upon exceeding
the maximum volume allowed on the display, the value
turns to zero and a new accumulation begins. This is
not accompanied by an alarm.
Stops pumping action and resets the VOL FED to zero.
NOTE: The pump will "beep" four times before clearing
the volume to alert users that the dial has been turned to
CLEAR VOL. The user must turn dial away from CLEAR
VOL (to any other setting) before the fourth beep to
avoid an unintentional loss of accumulated volume.
7
Tabla de contenido
loading

Productos relacionados para Abbott Patrol

Tabla de contenido