TROUBLESHOOTING
YELLOW or RED
LIGHT Illuminated.
Alarm:
CONTINUOUS
On red light only.
* Only applies to
SensO
models.
2
GREEN LIGHT with
YELLOW Light
Flashing.
Alarm:
RAPID
Beep....Beep...
Beep...Beep
PART ONE
1. Low oxygen purity.*
2. Kinked or blocked
tubing, cannula or
humidifier.*
3. Flowmeter set at
1/2 L/min. or less.*
4. Unit overheating
due to blocked air
intake.
5. Insufficient power at
outlet.
6. Internal repairs required.
1. Internal repairs required. 1. Call Service Provider.
LOW FLOW ALARM
1a. Kinked or blocked
tubing, cannula or
humidifier.
1b. Flowmeter set at
0.5 L/min. or less.
1. Clean or Replace filters.
2. Inspect for kinks or
blockages. Correct,
clean or replace item.
Once corrected, turn
power OFF for 60
seconds and then turn
power back ON.
3. Check flowmeter is set
to 1.0 L/min. or more.
Refer to Recommended
End-User Guidelines in
the Operating
Instructions of this
manual.
4.a.Remove and clean
cabinet filters.
4b. Move concentrator at
least 3- inches from
walls, draperies, and
furniture.
5. Do not use extension
cords. Move to another
electrical outlet or
circuit.
6. Call Service Provider.
1a. Inspect for kinks or
1b. Reset flowmeter to
20
blockages. Correct,
clean or replace item.
Once corrected, turn
power OFF for 60
seconds and then turn
power back ON.
prescribed flowrate. If
prescribed flowrate is
below 0.5 L/min., use
the Pediatric
Flowmeter IRCPF16.