Fora V12 El Manual Del Propietario página 40

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Disinfecting Procedures
1. Take out one disinfecting wipe from the package and squeeze out any
excess liquid in order to prevent damage to the meter.
2. Wipe all meter's exterior surface display and buttons. Hold the meter
with the test strip slot pointing down and wipe the area around the test slot
but be careful not to allow excess liquid to get inside. Keep meter wet with
disinfection solution contained in the wipe for a minimum of 2 minutes for
Micro-Kill+™ wipes. Follow the instructions on the package label of disin-
fecting wipe.
3. Remove the wipe. Allow the meter surface to dry completely.
4. Discard the used wipes and never reuse them. Wash your hands thor-
oughly with soap and water after handling the meter, lancing device and test
strips to avoid contamination.
This device has been validated to withstand up to 5000 cleaning and
disinfection cycles using the recommended disinfecting wipe/towelette. The
tested number of cycles is estimated by 2 cleaning and disinfection cycles
per day over 5 years, the expected life of the device. The meter should be
replaced after the validated number of cleaning and disinfection cycles or
the warranty period, whichever comes first.
Improper system cleaning and disinfection may result in meter malfunction.
Stop using the meter if you see any signs of deterioration, for example, LCD
display cracks or becomes cloudy, buttons no longer function, or outer cas-
ing cracks. If you have any question. Please contact local customer Care
Line for a replacement meter if any of the signs of deterioration are noticed.
40

Capítulos

Tabla de contenido
loading

Tabla de contenido