Agilent Technologies TV 551 SEM Navigator Manual De Instrucciones página 240

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 51
17
Technical Information
RS 232/RS 485 Communication Description
N.
108
109
110
111
117
120
121
122
123
124
125
126
240/276
Read/
Data
Description
Write
Type
(write only in stop)
R/W
N
Baud rate
W
L
Pump life/ cycle time/
cycle number reset
R/W
L
Interlock type
(default = 1)
R/W
L
Analog output type:
output voltage signal
proportional to
frequency or power
R/W
N
Low Speed frequency
(Hz)
R/W
N
Rotational frequency
setting (Hz)
R/W
N
Maximum rotational
frequency in Hz (active
only in Stop condition)
R/W
L
Set vent valve on/off
(on = closed)
Reserved to Agilent service
R/W
L
Set the vent valve
operation
R/W
N
Vent valve opening delay
(expressed in 0.2 sec)
TV 551/701 SEM Navigator User Manual /
Admitted Values
1 = YES
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
To reset write '1'
Impulse = 0
Continuous = 1
0 = frequency
1 = power
(default = 0)
272 to "Maximum rotational
frequency" (win 121)
(default = 476)
272 to 714 (default = 714)
272 to 714 (default = 714)
On = 1
Off = 0
(default = 1)
Automatic = 0 (see note 1.)
On command = 1 (see note 2.)
0 to 65535
(corresponding to 0 to 13107
sec)
87-900-958-01(G)
Tabla de contenido
loading

Tabla de contenido