Combics YDO02C-...: Installing in the Indicator
Installation
Installation of the interface module in the Combics indicator (including installation of the cable gland and attachment of the connection
cables to the screw terminal block) is only required if the customer did not order the respective data output port to be fitted in the factory.
Installation option as accessory
of the optional UniCOM interface
Interface module (RS-232)
Interface module (RS-485 and RS-485), electrically isolated
Electrically isolated digital I/Os, 5 outputs and
5 inputs, freely configurable
Analog current output, 0 – 20 mA, 4 – 20 mA, 0 – 10 V, 16 bit
Profibus DP interface module
Ethernet interface module
1
) suitable for use in zones 2 + 22
Notes:
§ The interface module should only be installed by a trained and authorized Sartorius technician.
§ IP69-K protection: Use the appropriate connecting cable with a cable gland designed for the interface module in question
(see "Accessories").
!
Disconnect the equipment from the electricity supply prior to commencing work.
Improper installation invalidates the applicable warranty.
!
Perform all installation work that affects the IP69K protection rating with extreme care.
!
The cable gland (IP69K protection) for connecting the interface to the indicator is sealed with a protective cap.
Please use extreme caution when performing any work on the cable gland.
1
)
§ Remove the front panel: Remove the 10 screws from the front plate.
Installing the UniCOM Board:
§ Unscrew the 2 fixing screws
CAISL1,
CAISL2/3,
CAW1P
CAW2P, CAW3P
for installation for installation
in IP44 version in IP44 version
1
)
CAIS1/2/3,
CAW1S/2S/3S, CAH1/3
for installation
in IP69K version
YDO02C-232
YDO02C-485
YDO02C-DIO
YDO02C-AO
YDO02C-DP
YDO02C-ETH
5 5 5