Compressor Cleaning; Filter Change; Maintenance - DeVilbiss PulmoNeb 3655DX Manual De Instrucciones

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

EN
3.
Keep the outer surface of the tubing dust-free by wiping regularly. Nebulizer tubing
does not have to be washed because only filtered air passes through it.
NOTE
– The Disposable Nebulizer will last at least 15 days and possibly longer,
depending on usage. Proper cleaning will also help extend the life of the
nebulizer. Because it is disposable, we recommend that an extra nebulizer be
kept on hand at all times. The Reusable Nebulizer is dishwasher safe and may
be cleaned and reused for up to one year.

COMPRESSOR CLEANING

1.
With power switch in the "
2.
Wipe outside of the compressor cabinet with a clean damp cloth every few days
to keep dust-free.
Do not submerge in water; doing so will result in compressor damage.

FILTER CHANGE

1.
Filter should be changed every 6 months or sooner if filter discolors.
2.
Remove filter (Fig. A) by grasping it firmly and pulling out of the unit. Discard old filter.
3.
Replace with a new DeVilbiss filter (part #3655D-601). Additional filters should be
purchased from your DeVilbiss provider.
4.
Push new filter into position.
CAUTION– Do not reuse or wash filter or substitute any other material such as
cotton for a DeVilbiss air-inlet filter. It will result in compressor damage.

MAINTENANCE

All maintenance must be performed by a qualified DeVilbiss provider or authorized service
center.
Electric shock hazard. Do not remove compressor cabinet. All disassembly
and maintenance must be done by a qualified DeVilbiss provider.
A-3655X
10
CLEANING & MAINTENANCE
" position, unplug power cord from wall outlet.
� DANGER
� DANGER
Tabla de contenido
loading

Este manual también es adecuado para:

Pulmoneb 3655txPulmoneb 3655ixPulmoneb 3655kxPulmoneb 3655px

Tabla de contenido