Technical Data; Operating Messages - Aqualytic AL 110 Manual De Instrucciones

Ocultar thumbs Ver también para AL 110:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 140

Technical Data

GB
Instrument
Light source:
Wavelength accuracy
Photometric accuracy*
Photometric resolution
Power supply
Operating time
Auto-OFF
Display
Storage
Serial Interface
Bluetooth
®
Modul:
Time
Calibration
Dimensions
Weight
Ambient conditions
Waterproof
CE
*measured with standard solutions
To ensure maximum accuracy of test results, always use the reagent systems supplied by
the instrument manufacturer.
Lo
btLo
70
Technical Data
three
wavelength, automatic wavelength selection,
direct reading colorimeter
LEDs,
interference filters (IF) and photosensor in transparent
cell chamber. Wavelength specifications of the IF:
430 nm
530 nm
560 nm
± 1 nm
3% FS (T = 20° C – 25° C)
0.01 A
4 batteries (AAA/ LR 03)
17hr operating time or 5000 measurements in constant
use when display backlight and Bluetooth
are off.
automatic switch off
20 minutes after last keypress
backlit LCD (on keypress)
internal ring memory for 125 data sets
Bluetooth
for the transfer of stored measurement data
®
Specifications:
Modul: BLE113-A
Bluetooth
4.0 LE
®
FCC ID: QOQBT113
IC: 5123A-BGTBLE113
real time clock und date
user and factory calibration
resetting to factory calibration possible
155 x 75 x 35 mm (LxWxH)
approx. 260 g (incl. batteries)
temperature: 5 – 40 °C
rel. humidity: 30–90 % (non-condensing)
floating; as defined in IP 68 (1 hour at 0.1 meter)
Certificate for Declaration of CE-Conformity
at www.aqualytic.de

Operating messages

Measuring range exceeded or excessive turbidity.
Result below the lowest limit of the measuring range.
Replace batteries, no further tests possible.
Battery capacity is too low for the display backlight;
measurement is still possible.
= 5 nm
= 5 nm
= 5 nm
data transfer
®
AL110_2 03/2018
Tabla de contenido
loading

Tabla de contenido