Specifications - Laerdal CPRmeter 2 Manual Del Usuario

Ocultar thumbs Ver también para CPRmeter 2:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Specifications

CPR Targets
Category
Compression Depth Target
Compression Release Target Force
Compression Rate Target
CPRmeter 2 [REF 801-002xx]
The CPRmeter 2 meets the performance requirements of
IEC 60601-1, 3rd edition.
Category
Dimensions
Weight
Batteries
Temperature
The time required for CPRmeter 2 to warm from the minimum storage
temperature between uses until it is ready for intended use is minimum
15 minutes at room temperature.
The time required for CPRmeter 2 to cool from the maximum storage
temperature between uses until it is ready for intended use is minimum
15 minutes at room temperature.
Sealing
Relative Humidity
Atmospheric
Pressure (Atm.p.)
Specification
> 50 mm ( 2")
Depth accuracy: ±10%
< 2.5 kg (5.5 lbs)
Force accuracy: +1.5 kg to - 2.0 kg
(+3.3 lbs to - 4.4 lbs)
100 to 120/min ± 3/min
Specification
153 mm x 64 mm x 25 mm (6.0" x 2.5" x 1.0")
163 g (5.7 oz) (excluding batteries)
2 x 1.5V Size AAA (LR03)
Transport and Storage: -20 °C to 70 °C (-4 °F to 158 °F)
Operation: 0 °C to 50 °C (32 °F to 122 °F)
Transient temperature: -20 °C to 70 °C
(- 4 °F to 158 °F)
Meets IP55 rating as defined by standard ISO/IEC 60529
Transport: 5% to 95%
Storage: 5% to 75%
Operation: 5% to 95%
Transport, Storage and Operation:
1060 to 617 mbar (1060 to 617 hPa)
26