The device is intended for use in the electromagnetic environment specified below. The customer or the user of device should assure that it is used in such an
environment.
Immunity test
IEC 60601test level
Conducted RF
3V
IEC 61000-4-6
0.15 MHz to 80MHz
6 V in ISM and
amateur radio bands
between 0.15 MHz
and 80 MHz
Radiated RF
10V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
56
EN
Compliance
Electromagnetic environment - guidance
level
Not applicable
Portable and mobile RF communications equipment should be used no closer to any
part of device, than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = 1,2
150 kHz to 80 MHz
d = 1,2
80 MHz to 800 MHz
d = 2,3
80 MHz to 2,7 GHz
10 V/m
where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
should be less than the compliance level in each frequency range
a
Interference may occur in the vicinity of equipment marked with the following symbol: