Setting Up the Transducer
1. Turn the transducer power supply on. The instrument will go through a
pre-programmed power-up sequence. The internal display will
sequence through the current settings: Calibration date "CALdAtE",
"=", month and year "03.1995"(March 1995), measurement units
"mEtErS"(m/s) or "FEET"(fpm), "=", selected output type and range
"0-5V" and the time constant "tc = 1.00" in seconds. The display will
then show velocity readings.
2. Push the SET button to enter setup mode. When setup mode is entered,
the transducer voltage/current output will drop below zero.
t
3. Use the
and
4. Use the SET button to advance to the next setup message.
Note:
The output signal is disabled during the 15 second warm-up period
and when the transducer is in setup mode.
Setup Message
SELECt UnitS
SELECt FULL
SCALE
SELECt OUtPUt
SELECt tc
AdJUSt ZErO
AdJUSt SPAn
s
buttons to select the desired setting.
FEEt (fpm) or mEtErS (m/s)
Select full scale of the velocity range
Select Output Type: 0-5V, 0-10V, 1-5V, 2-10V,
4-20mA, 0-20mA
Select Time Constant: 0.05 to 10 seconds (see page 11)
Adjusts voltage/current output signal at zero velocity
(see page 13)
Adjusts the voltage/current output by ±15% (see
page 12)
Available Settings
8