ADDITIONAL INFORMATION
1. Before cleaning your unit ALWAYS disconnect from the power supply and allow to cool
below 50° C.
Your calibrator can be cleaned by wiping with a damp soapy cloth. Care should be exercised
to prevent water from running inside the unit. Do not use abrasive cleaners.
2. The controller is factory preset. See the Controller book for further details.
For safety reasons, most of the controller functions are locked. Please contact your dealer
for further information
THE CONTROLLER MUST NOT BE SET TO CONTROL ABOVE 350°C OR DAMAGE TO
THE UNIT MAY RESULT.
3. If the over-temperature cutout light comes on on the front panel, a fault has occured and the
cutout must reset before continuing. To reset the cutout press the red button on the base of
the calibrator.
Repeated operation of the cutout indicates a serious fault: you may need to return the unit
to your supplier for repair.

Replacement Parts

The following parts may be obtained from your Techne dealer if replacements or alternatives
are required:
Techne Part Number Description
FCABLEUK
FCABLEEU
FCABLEUS
Mains cable with 13amp UK plug (5 amp fuse)
Mains cable with R/A Schuko plug
Mains cable with 3 pin US plug