• Use only approved thermal probe cable P/N D-0000404.
• External thermal probe is calibrated with the base unit. If thermal probe
is changed, then use FieldLab Desktop to update the serial number and
calibration coefficients of the new thermal probe.
Instructions for Cleaning or Decontamination
• Clean enclosure with mild water-based cleaner. Do not use aggressive
solvent.
• Flush pressure sensor or thermal probe with alcohol based cleaner (or other
cleaner compatible with 316 Stainless steel) if toxic or other hazardous
substance is used.
Button Functions
Power
Hi/Low/
Reset
Menu
Zero
Pressure
Backlight
• Press and hold to turn on.
• When LC20 is on, press and hold to turn off.
Note: LC20 will turn off automatically based on setting in
FieldLab Desktop. Default time is 10 minutes, but can be
changed using FieldLab Desktop software.
• Press button once to display highest pressure or
temperature reached.
• Press button again to display lowest pressure or
temperature reached.
• Press and hold button again to clear high and low values.
• Hi/Low/Reset applies to the Active Sensor. If there are 2
sensors, then change sensors to use Hi/Low/Reset for
the other sensor.
• If 2 sensors, quick press to change sensors.
• Long press initiates MENU and follow prompts.
• With no pressure applied, press and hold button to
zero reading.
Note: If pressure is greater than 10% of gauge pressure
range, then a message to vent the pressure will appear.
• Press button to illuminate backlight.
Note: Light will turn off after a set time. Both time and
brightness can be adjusted in FieldLab Desktop software.
25