E6 CHIP ERROR !
E7 MISSING TRAY !
E8 TRAY POSITION ERROR !
E9 WRONG TRAY !
E10 LIGHT BARRIER ERROR !
E11 MOTOR STEP ERROR !
E12 OPTICS ERROR !
CLOSE PRINTER COVER
NO PAPER IN PRINTER
E14 INTERFACE ERROR !
E17 INVALID PASSWORD!
48
Cause: The program chip on the right of the analyzer underneath the printer cov-
er (see Figure 18) is missing, is not making contact, is defective or contains an old
soft ware version.
Action: Switch off the Urisys 1100 system. Insert the program chip and switch the
instrument on again. If "CHIP ERROR" appears again, the instrument is defective.
Cause: No tray inserted or tray not inserted far enough to be engaged by the
motor.
Action: Insert the tray correctly (see Section 4). Press the START button.
Cause: The positioning hole in the tray (see Figure 17) is soiled or still wet after
cleaning; the retaining bar is open while the tray is advancing or the retaining bar
mechanism is fouled with urinary deposit and blocked. (See Section 7.1)
Action: Clean, blow through or dry the positioning hole (using a lint-free cloth) to
ensure that it is completely patent. Remove urinary deposits, if there are any, in-
cluding those on the underside of the tray. Insert the tray again and press the
START button. Ensure that the retaining bar is down and locked into place while
the reading is taking place. If the error message appears again, use the spare
tray.
Cause: The test strip tray used is not the one for the programmed test strip type,
or the grey reference pad is missing from the tray, or there is an instrument error.
Action: Press the START button. The strip type menu is displayed. The strip type
must match the tray type (see Section 11). Use the correct tray, or order a new
one. If the error message appears again, the instrument is defective.
Cause: The light barrier used to control the position of the test strip tray is defec-
tive or the tray transport is blocked.
Action: Pull out the tray and return it to the start position. Press the START but-
ton. If the error message appears again, the instrument is defective.
Cause: The stepping of the motor is out of tolerance or the advance of the tray is
blocked. This may be due to
- soiling on or between the cogs
- worn or broken cogs
- defective motor.
Action: Carefully clean the tray. Remove any urinary deposits, including those on
the underside of the tray and the cogs. If the tray is damaged insert the spare
tray. Press the START button. If the error message appears again, the instrument
is defective.
Cause: The reference pad is missing from the tray, or an LED or the phototran-
sistor is defective.
Action: Attach the reference pad or use the spare tray. Press the START button. If
the error message appears again, the instrument is defective.
Cause: The printer cover is open.
Action: Close printer cover.
Cause: No paper has been inserted or roll is finished.
Action: Insert new roll of paper and close printer cover.
After elimination of printer errors the results can be printed from the instrument's
memory using the "Print" function.
Cause: Fault in data transfer to PC or host in bidirectional or ASTM mode.
Action: Check the data cable. Verify that the PC or host is ready to receive data.
Use the "Send" function to transfer data or press "Home" to resume Ready-to-
Measure status.
Cause: The entered password doesn't match.
Action: Enter a correct password.