Recommended Use Of Control Solution - ReliOn Prime SYSTEM Manual De Instrucciones Para El Usuario

Sistema de monitoreo de glucosa en sangre
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

It is recommended to use Control Solution:
When you open a new bottle of test strips.
Whenever you suspect the meter or test strips may not be functioning properly.
If test results appear to be abnormally high or low or are not consistent
with clinical symptoms.
If the test strip bottle has been left open or has been exposed to light,
temperatures below 39°F(4°C) or above 86°F(30°C), or humidity
levels above 80%.
To check your technique.
When the ReliOn® Prime Meter has been dropped or stored below freezing
or above 122°F(50°C).
Each time the batteries are changed.