FEATURES
IP65 water resistant
Selectable temperature units, °C or °F.
Data hold function
Maximum and minimum reading memory
Stainless steel probe
SPECIFICATION
Temperature Range
Display Accuracy
Display Resolution
Automatic power off
Battery
Water resistant rating
Probe size
Ambient temperature
OPERATION
Power ON/OFF
1. Press [ON-OFF] to turn on.
2. Press and hold for approximately 3 seconds to turn off.
Select Temperature Unit
Press and hold [°C/°F] for 3 seconds and release to change the temperature unit.
Data Hold
1. Press the [HOLD] button once to retain display reading and the °C or °F icon will be flashing.
2. Press the button again will release the hold function and return to normal display.
Maximum/Minimum Reading Memory
1. Press the [MAX/MIN] to display maximum reading.
2. Press the button again to display minimum reading.
3. Press the button once more to display normally.
4. To reset the memory, switch the thermometer off then on.
* Reset the memory once before taking new readings.
NOTE
Clean the probe immediately after each measurement to avoid cross contamination.
When replacing battery, ensure all screws are tighten fully, otherwise it will affect the water resistant
function.
When the display becomes dim replace the battery.
If there is any abnormal display or operation, switch off the thermometer once. All user settings will
be reset to ex-factory settings.
SYMBOLS
Symbol
1) Sensor open circuit
LLL.L
2) The reading is out of lower limit (-50°C)
1) Sensor short circuit
HHH.H
2) The reading is out of upper limit (200°C)
-50°C to 200°C /-58°F to 392°F
±1°C between -19.9°C to 119.9°C / ±1.8°F between -3.8°F to 247.8°F
0.1° over a full range
After idle for 10 minutes
1.5volt, 1 x LR44 or equivalent (included)
IP65
3.5 (Ø) x 122(L)mm
0 to 50°C /32°F to 122°F
Description
Action required
1) Return the thermometer for repair
2) Keep the measurement above lower limit
1) Return the thermometer for repair
2) Keep the measurement below upper limit