3. To Disinfect: Using fresh wipes, make sure that all
outside surfaces of the lancing device remain wet for
2 minutes.
Super Sani-Cloths may be purchased at the following places:
• Amazon.com,
• Officedepot.com or visit your local store,
• Walmart.com
4. Let lancing device air dry thoroughly
before using to test. Replace End
Cap. Gently pull back Arming
Barrel and press the Trigger Button.
A click will be heard if the lancing
device is functioning properly.
5. Wash hands thoroughly again after
handling the lancing device.
Stop using the lancing device and call Customer Care for
assistance at 1-800-803-6025, Monday - Friday, 8AM-8PM
EST, if:
• Markings on lancing device are coming off,
• Trigger Button hard to push,
• End Cap does not go back on,
• Arming Barrel does not click when gently pulled back.
Note: Other disinfectants have not been tested. The effect of other disinfectants
used interchangeably has not been tested with the lancing device. Use of dis-
infectants other than Super Sani-Cloth Wipes may damage lancing device.
Note: Super Sani-Cloth Wipes have been tested on the lancing device for
a total of 1095 cleaning and disinfecting cycles, which is equal to
cleaning and disinfecting the lancing device once per day for a 3 year
period. Lancing device may be cleaned and disinfected up to 7 cycles
per week for 3 years.
42