Safe For Use In Conformance With Iec Regulations - B.Braun Aesculap GN200 Instrucciones De Manejo

Ocultar thumbs Ver también para Aesculap GN200:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 58
Aesculap
Lektrafuse HF Generator GN200
1.1

Safe for use in conformance with IEC regulations

The operating environment
Risk of death by electric shock!
Do not open the product.
DANGER
Only ever connect the product to power mains
with equipment grounding conductor.
Danger of injury from burns and/or explosion from
flammable gases! Sparks can occur even if the HF
surgical device is used properly and according to its
WARNING
intended use.
Do not use the device in explosion-hazard zones.
When operating in the head or thoracic region,
avoid using combustible anesthetics and accel-
erating gases (e.g. nitrous oxide or oxygen) or,
when using such substances, ensure they are
extracted from the region of operation.
If possible, use incombustible cleaning and dis-
infecting agents.
If combustible cleaning and disinfecting agents
or solvents have to be used: Verify that such
agents have evaporated prior to commencing HF
surgery.
Ensure that no combustible fluids accumulate
under the patient or in body cavities (e.g. the
vagina). Wipe off all fluids before using the HF
surgical device.
Ensure the absence of any endogenous, combus-
tible gases.
Check that oxygen-soaked materials (e.g.
absorbent cotton or mull) are kept at a safe dis-
tance from the HF field, so that they cannot
ignite.
Risk of interference with other devices!
Even during normal use, the HF surgical device cre-
CAUTION
ates electromagnetic fields that can interfere with
other devices.
Check that no electronic devices that could be
subject to interference by electromagnetic
fields are set up in the vicinity of the HF surgical
device.
Restriction to view and/or side-effects due to the
development of steam/smoke during HF surgery!
If necessary, use smoke suction.
CAUTION
Ensure that the device does not come in direct contact with the patient
or in the sterile area respectively.
Ensure that the user does not come into direct contact with the patient
and the HF device at the same time.
4
®
Patient safety
Danger due to inadequate preparation or faults in
the HF surgical device!
Check that the HF surgical device is in perfect
DANGER
working order.
Ensure that neither the foot switch nor the hand
switch has been penetrated by conductive fluids
(e.g. blood, amniotic fluid).
Ensure there is no short circuit in the foot or
hand switch cables.
Risk of burns suffered by the patient due to inad-
vertent activation of the HF surgical device!
DANGER
In case of any inadvertent activation of the HF
surgical device, switch off the device immedi-
ately at the OFF/ON switch.
Always exercise particular care when operating
the foot switch.
Risk of injury to the patient due to uncontrolled
rise of the HF output voltage due to some fault in
the HF surgical device!
DANGER
Stop using the HF surgical device as soon as it
shows even the slightest anomaly.
Risk of injury to patients/users due to defective
power cord or missing protective ground connec-
tions!
WARNING
Check the mains power cord/protective ground
connections.
Danger of injuries due to muscle contraction,
caused by stimulation of the nerves and muscles!
WARNING
Work with particular care on sensitive struc-
tures.