Sc-600-Eur Signal Clamp - Beha-Amprobe UAT-600-EUR Serie Manual De Usario

Ocultar thumbs Ver también para UAT-600-EUR Serie:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 116
Agency approval
Safety compliance
Electromagnetic Compatibility
Size (H x W x L)
Weight

SC-600-EUR Signal Clamp

Operating voltage & current
Operating frequency
Signal voltage
Output (nominal)
Operating temperature and
humidity
Storage temperature and
humidity
Operating altitude
Pollution degree
IP-rating
Drop proof
Measurement Category
Agency approval
Safety compliance
Electromagnetic Compatibility
Size (H x W x L)
Weight
IEC 61010-1
CSA/UL 61010-1
IEC 61326-1 Korea (KCC): Class A Equipment (Industrial Broadcasting &
Communication Equipment)
[1]
This product meets requirements for industrial (Class A) electromagnetic
wave equipment and the seller or user should take notice of it. This
equipment is intended for use in business environments and is not to be
used in homes.
Approx. 355 x 230 x 120 mm (14 x 9 x 4.7 in)
Approx. 1.9 kg (4.2 lb) (batteries installed)
0 to 600 V, 100 A max.
33 kHz (32,768 Hz) and 8 kHz (8,192 Hz)
23 V rms at 8 kHz
30 V rms at 33 kHz
-20 °C to 50 °C (-4 °F to 122 °F), ≤ 90 % RH
-40 °C to 60 °C (-40 °F to 140 °F), ≤90% RH
< 2000 m (< 6561 ft)
2
IP54
1 m (3.28 ft)
CAT IV 600 V
IEC 61010-1, IEC 61010-2-032
CSA/UL 61010-1, CSA/UL 61010-2-032
IEC 61326-1 Korea (KCC): Class A Equipment (Industrial Broadcasting &
Communication Equipment)
[1]
This product meets requirements for industrial (Class A) electromagnetic
wave equipment and the seller or user should take notice of it. This
equipment is intended for use in business environments and is not to be
used in homes.
Approx. 295 x 180 x 37 mm (11.6 x 7.1 x 1.4 in)
Approx. 0.85 kg (1.9 lb)