Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer
or the user of the device should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic
discharge
± 6 kV contact
(ESD) IEC
± 8 kV air
61000-4-2
± 2 kV for power
Electrical fast
supply lines
transient/burst
± 1 kV for input/
IEC 61000-4-4
output
Surge IEC
± 1kV line(s) to line(s)
61000-4-5
± 2 kV line(s) to earth
< 5% U
in U
Voltage dips,
short
40% U
interruptions
U
and voltage
variations on
70% U
power supply
U
input lines IEC
61000-4-11
< 5% U
in U
Power
frequency (50/
60Hz)
3 A/m
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
46/50
(> 95% dip
T
) for 0,5 cycle
T
(60% dip in
T
) for 5 cycles
T
(30% dip in
T
) for 25 cycles
T
(> 95% dip
T
) for 5 sec
T
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/
output
1 kV line(s) to line(s)
± 2 kV line(s) to earth
< 5% U
(> 95% dip
T
in U
) for 0,5 cycle
T
40% U
(60% dip in
T
U
) for 5 cycles
T
70% U
(30% dip in
T
U
) for 25 cycles
T
< 5% U
(> 95% dip
T
in U
) for 5 sec
T
3 A/m
FISDR - JAP / F1902124.ES / 07 / 2011 - updated 10/2016
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
If the user of the device
requires continued operation
during power mains
interruptions, it is
recommended that the
device be powered from an
uninterruptible power supply
or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.