4. If the average established in point 2 (page 90) differs from the dis-
play, the value must be changed step by step, pressing the display
switch each time.
To change the direction , the display switch must be pressed for
2 – 4 seconds.
1.00
1.01
SI
SI
Nm
ft·lbf
Nm
92
Rekalibrierung_4
IMP
SI
ft·lbf
Nm
ft·lbf
0.99
Rekalibrierung_5
SI
ft·lbf
Nm
ft·lbf
up
down
SI
SI
Nm
ft·lbf
Nm
5. Once the lower value is correctly set, hold down the display switch
for more than 8 seconds in order to change to Teach Mode for the
upper value. You can see that Teach Mode has been reached when
the display alternates between «CAH» (calibration high) and the
display value.
Please note: If you do not touch the switch, the device will remain
in this mode for 180 seconds before the electronic system reverts
to normal.
5.00
caH
8 sec.
8 Sek.
SI
SI
Nm
ft·lbf
Nm
SI
ft·lbf
Nm
ft·lbf
up
down
ft·lbf
93