Operation of Stand M LED with
Transilluminator Brightfield
Press the lower knob (Fig. 4/2) to switch on
the transmitted light illuminator and adjust it
by turning the illumination intensity control.
Toggle between brightfield (BF) and all-sided
darkfield (DF) illumination by moving the lever
(Fig. 4/3).
Press/turn
the
button (Fig. 4/1) to add a variable
amount of reflected light to the set
transmitted light.
Detailed information on operating the
optional
reflected-light
K LED
can
Stemi 305 / 508 operating manual.
4.2
Operation of Stand M LED with
Transillumination Unit
Press the lower knob (Fig. 5/2) to switch on
the transmitted light illuminator and adjust it
by turning the illumination intensity control.
To adjust the transmitted light brightfield
illumination, slide the rotary / slider button
(Fig. 5/3) to the rear position (BF).
Then adjust the tilting angle of the rotating
mirror by turning the button until the
specimen is brightly illuminated from below.
The rotating mirror has plain and
frosted mirror sides: use the plain side
for high-contrast brightfield, and the
diffuse (frosted) mirror for homogenous
brightfield.
To set lateral transmitted light darkfield
illumination, slide the control knob (Fig. 5/3)
to the front position (DF).
06/2016
OPERATION
upper
illumination
illuminators
be
found
in
the
435425-7144-008
Fig. 4
Operating the Transilluminator
Brightfield
Fig. 5
Operating the Transillumination Unit