The OtoRead™ OAE Test Instrument i™s designed to be a screening device for hearing loss.
Sensitivity and specificity of this type of device are based on the test characteristics defined by the
user, and may vary depending on environmental and operating conditions. The presence of
otoacoustic emissions suggests normal outer hair cell function, which in turn correlates to normal
hearing. However, a passing result using this instrument is not an indication that the full auditory
system is normal. Thus, a PASS result should not be allowed to override other indications that hearing
is not normal. A full audiologic evaluation should be administered if concerns about hearing sensitivity
persist. A REFER test result should not be assumed to be an indicator of a lack of auditory function;
however, it should be followed with full audiologic diagnostic testing.
on again by the normal procedure of holding the DOWN button.
The OtoRead™ Otoacoustic Emission Test System should be used only by those individuals trained
to perform the testing for which it has been designed. No person should attempt to use this instrument
without the necessary knowledge and training to understand how this equipment is to be properly
utilized and interpreted.
Never insert the probe tip into the ear canal without affixing an eartip.
1. Use only the disposable eartips designed for use with this instrument.
2. The eartips are disposable and for single use use only. Do not clean or reuse eartips.
3. Disposable 1.5v AA/UM-3/R6 Alkaline batteries are recommended for use with OtoRead.
If other batteries are used, they must be 1.6v AA Nickel Zinc rechargeable batteries. Care
must be taken to avoid bending the battery clips with incorrect sized batteries. Do not mix
battery types and do not mix old and new batteries.
OtoRead™ – Instructions for Use - EN
Intended Use
Under certain circumstances, especially in low-humidity environments,
common electrostatic discharge may cause this instrument to turn itself off.
This effect causes no permanent damage, and the instrument may be turned
Important Safety Instructions
Precautions
WARNING indicates a hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION, used with the safety alert symbol, indicates a hazardous
situation which, if not avoided, could result in minor or moderate
injury.
NOTICE is used to address practices not related to personal injury.