Technical Data
Ambient conditions
During operation:
Temperature
Atmospheric pressure
Relative humidity
Air velocity
During storage/transport:
Temperature
Atmospheric pressure
Relative humidity
Radiant heater
Radiant power
Distance between radiant
heater and resting surface
of bed, in cm
80
70
60
Light
Timer
Clearance between top edge of
radiant heater housing and ceiling
15 °C to 30 °C
900 hPa to 1100 hPa
0 to 90 %, no condensation
Max. 0.3 m/s
–20 °C to +60 °C
900 hPa to 1100 hPa
0 to 90 %, no condensation
Radiated power at
heater level 5,
2
in mW/cm
7
9
10
Halogen, 12 V / 35 W
0 to 99 minutes 59 seconds;
Start and stop times programmed in seconds;
Reset function;
Audible signal at 1 minute, 5 minutes and 10 minutes
>50 cm
Radiated power at
heater level 10,
2
in mW/cm
14
18
25