Kroeplin G0 Serie Manual De Instrucciones página 38

Palpadores electrónicos interiores
Ocultar thumbs Ver también para G0 Serie:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Programm: TOL
Programm: TOL
The measuring unit („mm" and „inch") and scale interval have to be
selected before setting the tolerance limits otherwise incorrect
measuring results may occur.
Configuration tolerance ranges: Example tolerance range 3 / Adjusting the tolerance limit – Example:
10,00 ±0,1mm in the „Absolute – Mode"
Activating the SETUP – Menu in the
1x long
„Absolute – Mode"
„tOL" will appear in the display
2x short
"1" for tolerance range 1 will appear in the display
1x long
The last selected status will appear in the display / Factory setting: 1
Switch to tolerance range "3"
2x short
„tOL LO" will appear in the display for 2s
1x long
The last saved lower tolerance value will appear.
38
Adjusting lower tolerance limit (9,90mm)
Tabla de contenido
loading

Este manual también es adecuado para:

G002G005G010

Tabla de contenido