6
UC32
Enzymatic Ultrasonic Cleaner
Concentrate in MeterDose
Dispensing Bottle (8 oz.)
UC34
Plaster & Stone Remover
Powder Concentrate in
Jar (1.85 lbs.)
UC38
Germicidal Ultrasonic Cleaning
Concentrate in MeterDose
Dispensing Bottle (8 oz.)
UC39
Cement Remover Concentrate
in MeterDose Dispensing
Bottle (16 oz.)
UC42
Germicidal Ultrasonic Cleaning
Concentrate in MeterDose
Dispensing Bottle (16 oz.)
XIII. Maintenance
A. Cleaning
All components should be cleaned
each day by wiping with a soft cloth
moistened with a commercially
available mineral deposit remover
followed by cleaning with isopropyl
alcohol or a mild disinfectant.
CAUTION: Do not use petroleum based
solvents, iodophors or phenolic based
products. (Iodophors and phenolics can
stain the surface of the unit.) Clean up all
liquid spills immediately. Rust deposits
on tank can be cleaned/removed with a
commercially available stainless steel
cleaner.
B. Inspecting the Power Cord
Periodically check the power cord for
damage and see that the connectors
at both ends of the cord are fully
seated.
C. Changing the Fuse
1. Remove the power cord from the
wall outlet first and then from the
power cord receptacle on the rear of
the unit.
2. The fuse compartment is located
directly next to the power cord
receptacle. To access, take a small,
flat-headed screwdriver to the tab on
the fuse holder. Apply slight pressure
to release the fuse holder (Fig. 12).
3. Gently remove the fuse from the
fuse holder.
4. Examine the fuse (Fig. 13). If blown,
replace with a fuse that matches the
unit's voltage and current (see
section XIV). If fuse appears to be in
good condition return the unit to your
dealer or Coltène/Whaledent.
5. Plug the power cord into the power
cord receptacle on the rear of the
236 ml
unit. Then plug the power cord into
the wall receptacle.
840 g
XIV. Unit Specifications
236 ml
Transducers: 2
UC125H unit
100 [100V 50/60 Hz, 337W (max.)]
115 [115V 50/60 Hz, 337W (max.)]
473 ml
230CE [230V 50/60 Hz, 337W (max.)]
473 ml
230UK [230V 50/60 Hz, 337W (max.)]
240 [240V 50/60 Hz, 337W (max.)]
Operating Frequency:
Tank Capacity:
Use Capacity:
Overall Dimensions:
Tank Dimensions:
Fuse Type:
Time Lag:
100-115V Heated units
230-240V Heated units
EMC: Class A
IEC 60417-5041
Caution, Hot Surface
ISO 3864, No B.3.
!
Caution: Refer to accompanying
documents
ISO 3864, No B.3.6
Caution: Risk of electric shock
XV. Warranty Information
Our products are carefully manufactured
to meet stringent quality assurance
requirements. Our products are
manufactured from new parts or new
and serviceable used parts. Regardless,
our warranty terms apply. This product
has been developed specifically for use
in dentistry and is intended to be
operated only by qualified dental
professionals in accordance with the
instructions contained in this guide.
However, notwithstanding anything
contained herein to the contrary, the user
shall at all times be solely responsible for
determining the suitability of the product
for the intended purpose and the method
of its use. Any guidance on application
technology offered by or on behalf of the
manufacturer, whether written, verbal or
by demonstration, shall not relieve the
dental professional of his/her obligation
to control the product and to make all
professional judgments regarding its use.
Our products are warranted in
52 kHz
accordance with the terms of a written
Certificate of Limited Warranty
1.25 gallon or 4.75 L
accompanying each product. Except for
1 gallon or 3.78 L
the warranties specifically set forth in the
Certificate of Limited Warranty,
16.5 x 10.5 x 13.8 in.
Coltène/Whaledent Inc. provides no
42 X 27 X 35 cm
warranties or guarantees of any kind
11.25 x 5.62 x 5 in.
covering the product, expressed or
28.6 x 14.3 x 12.7 cm
implied, including, without limitation, any
warranties as to merchantability or
5 x 20 mm Cartridge
fitness for a particular purpose. The
IEC 60127-2 Sheet 3
purchaser/user is referred to the Certificate of
Limited Warranty for all of the terms,
4 Amp
conditions and limitations of the warranty
2.5 Amp
covering this product. This Section of the
user manual is not intended to in any
way modify or add to the warranty
provided in the Certificate of Limited
Warranty.
Any claim for damage or breakage to the
product in transit should be made to the
carrier promptly upon discovery. C/W
does not warrant the product against
shipping damage.
®
BioSonic
UC125H