Setting Up Your Meter- Setting the Time, Volume and Date
You can enter Setting Mode by below two ways.
1. Reload battery
After removing the battery, please press the main button for several times until no signal on screen, then follow the
battery installation steps to load batteries. The
Rightest
button to exit the self test and enter Setting Mode.
2. With Battery inserted
®
Press the main button first to turn on the
Rightest
for 7 seconds (the meter will first show backlight, then be turned off during this period, please keep pressing) until you
hear a beep, indicating you have successfully entered Setting Mode. The display screen will show setting data.
NOTE
- When you keep pressing the main button for 2 seconds, the display on screen will show backlight. Keep
pressing the main button till display is off.
- When you keep pressing the main button for 4 seconds, the display on screen will turn off. Please don't care it.
Keep pressing the main button till entering setting data.
- Quick press the main button allows you to change setting. However, if you want to return to the time mode, you
need to go through all the settings first by pressing quickly the main button.
18
®
Blood Glucose Meter GM550 will do self test. Press the main
Blood Glucose Meter GM550. Then press and hold the main button
Setting Up Your Meter- Setting the Time, Volume and Date
1. Year setting
With the year format blinking, press the Left or Right button to adjust it. Then press the Main
button to confirm it. Then it will move to Month setting.
2. Month setting
With the month blinking, press the Left or Right button until the current month appears. Then
press the Main button to confirm it and move to Day setting.
3. Day setting
With the day blinking, press the Left or Right button until the current day appears. Then press the
Main button to confirm it and move to Time format setting.
1)
2)
3)
19