14. Function
Stationary use
14.1 Press the trigger button.
• Water flows.
☞
The trigger button remains in position so that the user can keep his eyes open with
both hands.
14.2 Press the red stop button to stop the water flow.
Mobile use
☞
The eye-wash fountain can be removed from the wall holder depending on the
length of the connection hose.
15. Maintenance and Servicing
The functionality of the eye-wash fountain must be checked every month. (Directives
for laboratories BG
As a preventative measure to reduce microbiological contamination, it is
recommended to change the water in the fitting by opening the valve in short
intervals (e .g every week).
Use suitable cleaning agents that are not harmful to the fittings in a proper manner
and rinse with water after use. High-pressure cleaners must not be used for
cleaning.
16. Replacing the Filter
Order no.:
70 - 1306 10 36
70 - 1316 10 36
70 - 1326 10 36
16.1 Close the isolating valve.
16.2 Loosen the union nut (a).
☞
Ensure correct installation position.
16.3 Replace the filter (b).
☞
Important!
The filter contains the connection seal.
16.4 Reassemble in the reverse order.
Order no.:
70 - 1336 10 36
16.5 Close the isolating valve.
16.6 Loosen the union nut (d).