Max Level; Peak Hold; Response Time; Calibration - Phonic PAA3 Manual Del Usuario

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Max Level

Maximum level your audio has reached will constantly be
shown on the LCD screen. The user should reset the max
level before each new measurement.
SETTING
MAX LEVEL
RESET
ESC
A 31
100 315 1K
Procedure:
Enter the SETTING menu, then onto the MAX LEVEL
1.
menu.
Highlight RESET and press ENTER button to reset the
2.
max level measurement (doing so will automatically
return you to the SETTING sub-menu). Press ESC
to return if it you do not want to reset.
A new max level display will be shown in the MAX
3.
column three seconds after returning to the RTA
operation.
Quick Refresh: While on the RTA screen, simply press
the left and right buttons simultaneously to refresh the
maximum level.

Peak Hold

Peak hold allows the user in RTA mode to constantly display
the peak values of all measured frequencies, individually
or collectively.
Procedure:
Enter the SETTING menu, then select the PEAK
1.
HOLD option.
Highlight and select ON to activate peak hold display.
2.
Highlight and select OFF to cancel the peak hold
3.
display function.
Press ESC to return to the SETTING menu.
4.
ATTENTION: When saving the memory with "PEAK HOLD"
on, the peak level will only be displayed when the user
recalls that memory when the PEAK HOLD setting is on.
By setting the PEAK HOLD off, the user may then view the
levels from all frequencies, individually or collectively.
12
3.15K
10K

Response Time

Response time settings will vary according to the user's
measurement goals.
35 ms: Extremely Fast (for explosive sound)
125 ms (F): Fast
250 ms (M): Medium
1 sec (S): Slow
SETTING
RESP TIME
35 ms
125 ms (F)
250 ms
250 ms (M)
1 se
c (S)
ESC
A 31
100 315 1K
Procedure:
Go to the SETTING menu, then enter the RESP TIME
1.
menu.
Select an appropriate response time (a tick will appear
2.
next to your selection), or ESC to cancel.

Calibration

You should never actually need to calibrate your PAA3. If
measurement data or operation of the unit is abnormal,
however, it may be necessary to. Anyone can calibrate
the PAA3 and regain accurate sound pressure level
measurement by using a sound level calibrator with 1/2"
diameter adapter that sends out a 1 kHz tone. A B&K TYPE
4231 sound level calibrator is suggested.
Procedure:
First, turn the PEAK HOLD OFF and set the RESP
1.
TIME (response time) to 250 ms.
Place a sound level calibrator with a microphone
2.
connector of a 1/2" diameter close to the PAA3's
built-in mic.
Enter the SETTING menu and activate the function.
3.
Adjust the level measured from the SPL calibrator by
4.
pressing UP/DOWN buttons until the level is equal
to that of the sound level calibrator (typically 94 dB).
Pressing the UP button each time will increase the
value by 0.1dB; pressing the DOWN button each time
will decrease by 0.1 dB.
Press the ENTER button to complete the calibration
5.
and to return to RTA display.
ATTENTION: If you want to cancel calibration, simply press
the ENTER button to exit before making any adjustment.
To resort the default calibration, simply adjust the offset
value to 0.0 dB.