Inogen One® G4 Column Change Procedure (continued)
9. Insert column assembly into the Inogen One® G4
concentrator. Do not leave the column ends exposed;
column assembly should be inserted into the Ino-
gen One® G4 as soon as the dust caps have been
removed.
10. P ush the column assembly into the device such that
the columns are fully seated into the Inogen One®
G4 concentrator. The spring loaded latch button
should fully return to the closed position.
11. Connect the AC power supply cord to the Inogen
One® G4 and plug the power supply AC cord into an
electrical outlet. Do Not Power on the Inogen One®
G4 concentrator.
12. Press and hold the plus (+) and (-) minus button for 5
seconds. The screen will display "sieve reset"
message. Release button once message is displayed
on screen.
13. Press the alert
display "sieve reset success".
14. P ress the power button to turn on the Inogen
One® G4, and use normally.
WARNING Do not use any columns other than those specified in this user manual.
The use of non-specified columns may create a safety hazard and/or
!
C
impair equipment performance and will void your warranty.
button once and screen will
Cleaning, Care and Maintenance Chapter 6