4. Install a replacement fuse, Inogen RP#125 (BUSS MDA-12) and reassemble the tip.
Ensure the retainer ring is properly seated and tightened.
Cigarette Adapter Plug
CAUTION For continued protection against risk of fire, only use fuse specified.
Inogen One® G4 Column Change Procedure
NOTE
Column change instructions are only to be used when maintenance is
required and are not intended for practice purpose.
1. Turn off the Inogen One® G4 concentrator by pressing the power button to shut
down the device.
2. Remove the Inogen One® G4 concentrator from the carry case, if applicable.
3. Remove the battery from the Inogen One® G4 concentrator.
4. Place the Inogen One® G4 concentrator on its side so
that the underside is visible. The metal column as-
sembly can be seen on one side of the device.
Fuse
Retainer Tip
Cleaning, Care and Maintenance Chapter 6
Retainer
33