5.7  I nstrument self-test
The instrument self-test requires user interaction.
1
In the Setup menu, select "6. Instrument Self-test".
Selecting the menu item starts the self-test routine.
2
Press the function keys on the keypad one by one in any order.
The self-test result is displayed after a few seconds.
The meter returns to the system settings menu automatically.
The user needs to finish pressing all seven keys within two minutes, otherwise "Self-test
failed!" appears and the procedure has to be repeated.
If error messages repeatedly appear, contact METTLER TOLEDO Service.