SIMPLE LOGGER II
MODEL
Electrical
Inputs
Input connections
Input levels
Uncertainties (50 / 60 Hz)
Resolution
Maximum input level ***
Input impedance
Sample rate
Storage rate
Storage modes
Recording length
Memory
Communication
Power source **
Battery life
Mechanical
Dimensions
Weight (with batteries)
Case
Vibration
Shock
Drop
Environmental
Operating temperature
Storage temperature
Relative humidity
Altitude
Safety
Conformity
Level of protection
*See the end of chapter
BNC
0 to 1 V for use with sensor
delivering voltage output
0 to 10 mV: unspecified
10 to 50 mV:
± (0,5% of reading + 1 mV)
50 to 1000 mV:
± (0,5% of reading + 0,5 mV)
0,1 mV
5 Vrms ± 7,07 Vpeak
800 kΩ
Programmable from 125 ms to 1 day
Start/Stop, FIFO and Extended Recording Mode* (XRM™)
Programmable using DataView
240,000 measurements (512 kB). The recorded data is stored in non-
volatile memory retained even if the battery is low or removed.
USB 2.0 optically isolated
2 x 1.5 V AA (LR6) cell Alkaline batteries
100 hours to >45 days (dependent on storage rate/recording length)
136x70x32 mm (5.38x2.75x1.28")
IEC 68-2-6 (1.5 mm, 10 to 55 Hz)
IEC 68-2-27 (30 G)
- 10° C to + 50° C (14° F to 122° F)
- 20° C to + 60° C (- 4° F to + 140° F)
up to 85% at 35° C (95° F) without condensation
EN61010-1; 600 V CAT III; 300 V CAT IV; Pollution level 2
L562
Two
Two 4 mm safety banana jacks
0 to 600 V
0 to 5 V: unspecified
5 to 50 V:
± (0,5% of reading + 1 V)
50 to 600 V:
± (0,5% of reading + 0,5 V)
1,2 X 600 V
40 MΩ
64 samples/cycle
®
from 15 minutes to 8 weeks
180 g (6.4 oz)
UL94-V0
IEC 68-2-32 (1 m)
2000 m
IP 40
English
AC/DC
0,1 V
Page 46