5.7. Memory – displaying the last 14 measurements
The measured results are stored in the instrument until a new
measurement is carried out or the batteries are removed. With the
unit in the OFF position, press and hold the MEMORY button for at least
3-seconds. The screen will show the last reading along with time and date.
The MR symbol is displayed in the lower left. Release the button.The last
reading will be displayed. Each time you press the memory button an
earlier measurement will be displayed along with time and date.

5.8. Discontinuing a measurement

If it is necessary to interrupt a blood pressure
measurement for any reason (e.g the patient
feels unwell), the ON/OFF button can be pressed
at any time. The device then immediately lowers
the cuff pressure automatically.
Further information
Measurements should not occur soon after each other, since
the results will be inaccurate. Wait several minutes in a relaxed position,
sitting or lying, before you repeat a measurement.

6. Error Messages/Troubleshooting

If an error occurs during a measurement, the
measurement is discontinued and a corresponding
error code is displayed.
(example: error no. 1)
Error No.
ERR 1
ERR 2
14 readings
in memory
Possible cause(s) / Solutions
The systolic pressure was determined. The
tube may have loosened, or no pulse was
detected.
*Ensure cuff connections are tight with
proper cuff placement. See section (5.3)
Unnatural pressure impulses. Reason:
The arm was moved during the
measurement (Artefact).
Repeat measurement keeping arm still.
Last
reading
14
MR
MR
12