The button
display in the centre of the display (function: DC voltage/ current). In
the DATA LOG function, the bar graph centring function is not available.
Switch-over by briefly pressing a key. By pressing the
longer (2 seconds), the displayed value can be switched over from
20,000 digits to 4,000 digits. This can be reset by pressing the key
again (2 seconds) or by switching off the BENNING MM 11. This mode
is not available for the following ranges:
7.5/ 7.6 Resistance measurement 200 MΩ and 2 GΩ
7.7 Diode and continuity testing
7.8 Capacity measurement
7.9/ 7.10 Frequency and pulse duty ratio measurement
7.11/ 7.12 Temperature measurement
A switch-over to the lower displaying value might be necessary, if for
V AC and A (mA) AC measurements the frequency shall be displayed
simultaneously in the lower display 2. For this, the measured value (V;
A; mA) must be at least 30 % of the displayable range.
5.1.9
The button
Max-/ Peak Min" value (function: mV, V, mA, and A) with a simultane-
ous indication of "PH" in the display. Each new save of the "Peak Max-/
Peak Min" value is confirmed by an acoustic signal. The "Peak Max-/
Peak Min" value is retrieved by pressing the button
cated in the subdisplay  . A renewed operation of
switches back into a standard mode.
5.1.10 The function
values in the subdisplay . In the display  the symbol "AH" is shown
simultaneously. As soon as a measured value is input, the subdisplay
 is updated with the new value. Unstable measured values or noisy
signals cannot be stored via the
button switches back into measuring mode. In the following measuring
functions, the device switches over from the automatic to the manual
measuring range to store the measured values (
measurement, capacity measurement and frequency measurement.
5.1.11
The button (blue)  selects the secondary or tertiary functions of the
rotary switch setting:
Rotary switch setting:
VAC,mVAC
Ω (Messspannung: 3,3 V)
mAAC,AAC
Note:
-
The function LVΩ (resistance measurement) reduces the measuring
voltage to 0.6 V, so that semiconductor elements operated in conduc-
tion direction such as diodes or transistors do not open.
-
Function %DF (Duty Factor) describes the pulse duty ratio of periodic
signals:
DF [%] =
5.1.12 The button
lowest measured value as well as the mean value in the subdisplay
. The following values are indicated by indexing: "MAX" indicates the
highest measured value stored, "MIN" indicates the lowest measured
09/ 2007
 centres the zero point of the analogue bar graph
 (peak value storage) inputs and stores the "Peak
 (measured value store) stores stable display
Diode
Capacitor
Hz
°C
T1
T
 inputs and stores automatically the highest and
­ B ENNING­MM­11­
-function. A renewed press of the
Secondary function:
VDC,mVDC
LVΩ (reduced
measuringvoltage: 0,6 V)
acoustic continuity testing
mADC,ADC
% DF (Duty Factor),
pulse duty ratio
°F
key 6
 and indi-
-Button 
). Resistance
Tertiary function:
VAC+DC,mVAC+DC
mAAC+DC,AAC+DC
9