First Steps

Is everything in the package?
Confirm that your package includes:
Bottle-top dispenser Dispensette
tube with recirculation valve, recirculation tube (included only in recirculation valve models), mounting tool,
three bottle adapters, a performance certificate and this operating manual.
Nominal volume, ml
10

Assembly

Warning:
Wear protective clothing, eye protection and gloves! Follow all safety
instructions and observe limitations of use and operating limitations
(page 28-30).
1.
Check safety ring
Note:
The discharge valve and discharge tubes are labeled with the spring
material. The label 'Pt-Ir' or 'Ta' must match for each device. The
spring material is selected based on the application ('Recommended
application'; see table on page 30).
Before use in trace analysis, the instrument must be thoroughly clea-
ned first. See 'Cleaning for trace analysis' (see page 42).
2.
Mounting the
filling tube/ recir-
culation tube
32
S Trace Analysis, telescoping filling tube, discharge tube or discharge
®
Adapters for
bottle thread
GL 28/ S 28 (ETFE), GL 32 (ETFE),
GL 38 (ETFE), S 40 (PTFE)
Make sure that the safety ring is firmly
tightened (Fig. 1).
Adjust the length of the telescoping fill-
ing tube to the bottle height and attach
it. Center and attach the filling tube (part
with smaller diameter) carefully to avoid
damaging the nozzle. If a discharge tube
with a recirculation valve is used, the
optional recirculation tube must also be
installed. Insert it with the opening point-
ing outward (Fig. 2).
Filling tube
Length, mm
125-240
1
Opening for
recirculation tube
Olive for filling
tube
2