CE-Declaration of conformity
Safety instructions
Unpack
Heating
Spare parts list
Spare parts diagram
CE-Declaration of conformity
We declare under our sole responsibility that this product corre-
sponds to the regulations 2006/95/EC and 2004/108/EC and con-
forms with the standards or standardized documents DIN EN IEC
61010-1, -2-010 and DIN EN IEC 61326-1.
2009-03-18
Safety instructions
EN
For your protection
Page
• Read the operating instructions in full before starting up and
15
follow the safety instructions.
15
• Keep the operating instructions in a place where they can be
16
accessed by everyone.
16
• Ensure that only trained staff work with the appliance.
17
• Follow the safety instructions, guidelines, occupational health
and safety and accident prevention regulations.
• Wear your personal protective equipment in accordance with
the hazard category of the medium to be processed. Otherwise
there is a risk of splashing liquids.
• Set up the device in a spacious area on an even, stable, clean,
non-slip, dry and fireproof surface.
• Prior to each use, always check the device and accessories for
damage. Do not use damaged components.
• CAUTION! Only process and heat up media that has a flash point
21
higher than the adjusted safe temperature limit of the heating
22
bath that has been set.
23
The safe temperature limit of the heating bath must always be
24
set to at least 25 °C lower than the fire point of the media used.
24
• DANGER OF BEING BURNT! When in operation the upper cor-
24
ner of the bath can heat up to 180 °C.
25
• Prior to filling or emptying the heating bath, the device must be
56
switched off and disconnected from the power supply at the
58
plug.
• Only fill or empty the heating bath when it is cold.
• Empty the heating bath prior to transporting it.
• Never operate the heating bath without tempering medium.
• CAUTION! The preferred tempering medium in the heating bath
is water (up to approx. 80 °C). Low viscosity silicone oils (50 mPas)
with a flashpoint > 260 °C are also permitted. There is a risk of bur-
ning when using tempering media with lower flashpoints!
• When using as a water bath, the use of demineralised water is
recommended.
• Always observe the minimum fill level of one litre when using oil
as the tempering medium.
15