Memmert SNxx PLUS Serie Instrucciones De Manejo página 65

Tabla de contenido
T T echnical description according to EN 60601-1-2
The steriliser type S.. is intended for use in electromagnetic enviroment in which radiated RF disturbances
are controlled. The customer or of the steriliser type S.. can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the steriliser type S.. as recommended below, according to the maximum power of the
communications equipment.
Rated maximum output
power of trtansmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
d
distance
in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, whre
transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and poeple.
Description of terminal
LAN port
Warning!
D33342 | Fecha: 12/2019
Recommended separation distances between portable and mobile
RF communications equipment and the Memmert steriliser type S..
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d
P
= 1,2
0,12
0,38
1,2
3,8
12
P
is the maximum output power rating of the transmitter in watts (W) according to the
Listing of cables and maximum length of cables
The use of other cables may result in increased emissions or
decreased immunity of the Memmert steriliser type S..
m
80 MHz to 800 MHz
d
P
= 1,2
0,12
0,38
1,2
3,8
12
Type of cable
RJ45 CAT 6
800 Mhz to 2,5 GHz
d
P
= 2,3
0,23
0,73
2,3
7,3
23
Maximum length of cable
m
2
Anhang
65