Operation

Operation

Short Overview and Front Panel Pad

Functions
Feeding film
Before feeding the next film wait until the LED stops
flashing, but is continuously on. An acoustic signal
additionally indicates that the next film can be
inserted.
Power (ready to operate)
The LED is on as long as the unit is on.
Warming up
During warm-up the LED flashes; before feeding the
next film wait until the LED stops flashing, but is con-
tinuously on. The end of the warm-up phase is indi-
cated by an acoustic signal. If you try to insert a film
during warm-up, an interval sound will occur perma-
nently.
Bath heating failure
If the developer temperature is not reached within 30
minutes, the LED starts flashing rapidly and an inter-
val sound occurs permanently.
Important!
Safety function stops film transportation when cover is re-
moved. Therefore keep cover placed on the machine when
processing films.
16
LED
Symbol for film feed released
Symbol for film feed locked
ECOMAX