LumiThera VALEDA Manual Del Usuario página 55

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Electromagnetic Immunity for All Equipment and Systems
Valeda is intended for use in the electromagnetic environment specified below. The operator of
Valeda should ensure that it is used in such an environment.
Immunity
Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical Fast
Transient/Burst
IEC 61000-4-4
Surge Transient
IEC 61000-4-5
Voltage Dips and
Interruptions
IEC 61000-4-11
Power Frequency
Magnetic Field
IEC 61000-4-8
Compliance
Level
± 15 kV air
± 8 kV contact
2 kV at AC
power lines
2 kV line-to-earth
at AC power ports
1 kV line-to-line at
AC power ports
- 0%, 0.5 Cycle
- 0%, 1 Cycle
- 70%, 25 Cycles
- 0%, 250 Cycles
(interruption)
30 A/m; 50/60 Hz
Electromagnetic
Environment – Guidance
Floors should be made of
wood, concrete or ceramic
tile. If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Power frequency magnetic
fields should be that of a
typical commercial or
hospital environment.