The camera will move to the chosen treatment eye and will turn on to allow viewing and positioning
of the beam on the patient's eye.
Patient Preparation
1. Verify that the patient has removed glasses and contact lenses.
2. Position the patient's forehead against the forehead rest. If the patient's forehead is not
detected, the system will display "No patient detected."
3. Adjust the chin rest elevation to position the patient's eyes vertically in the centre of the light
aperture (Figure 9).
4. Adjust the patient's chin position on the chin rest as shown below (Figure 9).
5. Confirm the patient's hair is outside the light aperture.
6. Monitor the patient throughout treatment to ensure proper patient alignment with the system is
maintained.
FIGURE 8
FIGURE 9
10