7. Stop insufflation: Press the STOP function field.
The following values are displayed:
Actual pressure 0.0 [mmHg]
Gas consumption >0.0 [l]
8. Press the gas volume function field.
Gas consumption >0.0 [l]
The basic function test of the device is complete.
11.3

Testing the Pressure Sensors in High Flow Operating Mode

(1)
(4)
1. Set to High Flow operating mode.
2. Select a nominal gas flow rate of 1.0 l/min.
Do not press the START/STOP function field.
CAUTION!
Never use the syringe to extract gas from the device.
3. Connect a manometer (1) and an air-filled syringe (4) to the insufflation
tube connection (3)/adapter (2).
4. Use the syringe to generate a pressure of at least 10 mmHg, which registers
on the manometer.
Actual pressure display: 10 ±2 [mmHg]
5. Use the syringe to generate a pressure of at least 20 mmHg, which registers
on the manometer.
Actual pressure display: 20 ±2 [mmHg]
6. Use the syringe to generate a pressure of at least 30 mmHg, which registers
on the manometer.
Actual pressure display: 30 ±2 [mmHg]
(3)
(2)
Annual Inspection
EN
51