Tabla 19 Parámetros del bus del bloque de transductores (continuación)
Parameter
PROCESS_PRESSURE
SALINITY
Calibration
CALIBRATION_MODE
CAL_HIGH
CAL_ZERO
CAL_SETTLINGTIME
CAL_RH
CAL_PRESSURE
CAL_TEMP
Description
Sets the process pressure
Sets the salinity value
Sets the calibration mode
The slope of the Oxy sensor in
Ampere
The zero offset resulting from a
calibration in Ampere
Shows the settling time in
seconds
Shows the relative humidity in
%
Shows the pressure
Shows the temperature which
was used during calibration
Data
R/W Bytes
Range
type
0–9.999 BAR
0–999.9 kPa
R/W 4
float
0–145 PSI
Default: 1.013 BAR
0–45 g/kg
R/W 4
float
Default: 0 g/kg
0 = Sat
R/W 1
uns8
1 = Conc
30–110 Sensor A
R/W 4
float
220–525 Sensor B
Default: 60 E-9A
–2–2
R/W 4
float
Default: 0 E-9A
R
4
float
Default: 0 s
R
4
float
R
4
float
R
4
float