Battery Replacement; Fuse Replacement; Cleaning And Disinfection - HERSILL V7 Instrucciones De Uso

Equipo de aspiración médica
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

7.1.2. Battery replacement

(V7m, V7mx, V7plusb and V7plusb emergency)
1. Unplug the unit from the mains.
2. Withdraw the 6 screws from the back side.
3. Maintain the unit in vertical position and separate the both parts of the casing.
4. Take off the old battery terminals and recycle it pursuant the local rules.
5. Set up the new battery with its terminals in its correct polarity.
6. Close the casing and set up the 6 screws in the back-side.
7. Charge the battery for 5 hours and test its capacity (Chapter 7.1.1).

7.1.3. Fuse replacement

Unplug the unit from the mains
Replace the fuse with a new one of the same ratting. (See Technical
specifications)
12/24Vdc fuse:
turn left using a
flat head
screwdriver and
take the fuse
holder out

7.2. Cleaning and Disinfection

Disconnect the machine from any external mains before cleaning it. Use a
minimum quantity of water or other liquids to avoid an electric discharge.
Contaminated devices must be thoroughly cleaned prior to disinfection or
sterilization
Personal protective equipment must be worn when handling contaminated
equipment, particularly tubing, filter and collection jar (protective clothing,
disposable latex gloves, mask, eye protection)
If any liquid accidentally enters in the main unit or in the pump the unit may
be damaged. For this reason, we recommend not to use the aspirator itself
as a method to clean the collection jar (taking in water with disinfectants)
nor connecting the suctioning tubes in the opposite way (the safety device
would not work). The failure should invalidate the guarantee and the
technical specifications under the manufacturer's responsibility.
In the event that contaminated fluids have moved past the internal unit,
remove the tubing and the filter, clean the casing (7.2.1), and
decontaminate the collection bottles (7.2.2). Return the device for service.
Polycarbonate jars can be reused up to 40 times and Polysulfone jars up
to 100 times. Before each use, check that collection jars are not damaged.
To obtain additional information including questions on infection control procedures,
please contact Hersill or your dealer. Telephone: +34 91 616 41 11
ENGLISH
110-240Vac fuse:
Using a flat head
screwdriver, pry out
the plastic fuse holder
8