REV.9
that you wash your covermats more often – after every bath if possible.
6. IF I LEAVE MY BATTERY HAND CONTROL ON CHARGE WILL I DAMAGE
IT?
No, charge your battery hand control for 12 hours when you first receive your
bathlift. Then after every bath, leave the hand control on charge. You cannot
overcharge your battery hand control.
A fully charged battery will provide
approximately fifteen full lifts up and down. Only charge your battery handcontrol
on a hard flat surface at normal room temperature and do not cover.
7. HOW DO I CLEAN MY BATTERY HAND CONTROL?
To clean inside the foam of your battery hand control simply remove the two
screws at the back of the hand control and pull the foam away from the control
box. Clean as required and simply replace your battery hand control back inside
the foam and insert the screws as previous.
8. DO I NEED TO USE MY SIDEFLAP DIVERTERS?
(Grey/White plastic strips packed with every bathlift)
Yes, if you have handles on your bath, please ensure you attach the side flap
diverters to your bath handles. Failure to do this may cause an obstruction to the
mechanism and will affect your warranty. If the bathlifts side flaps become trapped
under your bath handles, stop, press the down button on your hand control, make
sure the side flap is no longer trapped under the handles and press the up button
on your hand control.
TECHNICAL DATA
Length of seat: 580 mm ( 22.8 inches)
Width of seat with extended seat side flaps: 686 mm ( 27 inches)
Width of seat with seat side flaps folded in: 390 mm ( 15.3 inches)
Height of seat in its lowest position: 90 mm ( 3.5 inches)
Height of seat in its highest position: 465 mm ( 18.3 inches)
Depth of seat measured from
Backrest to front seat edge: 490 mm ( 19.3 inches)
Height of backrest: 604 mm ( 23.8 inches)
Width of backrest: 340 mm ( 13.4 inches)
Carrying capacity: 170kgs (26.7 stone, 175lb)
Total Weight without battery: 13.64kg (30 lb)
Weight of seat: 7.70kg (16.94 lb)
Weight of sideflaps: 0.94kg (2.07 lb)
Weight of backrest: 5kg (11 lb)
Weight of battery hand controller: 670g
Duration of stroke (80 kg load): Up approx. 50sec, Down approx. 35sec
Operating voltage: 13.2V DC
Maximum current consumption: 5.8A
Nominal Input: 13.2V DC
Motor noise with no load : approx. 50dB
(according to DIN 3746)
Medical equipment type B
Class Protection: IP 66
Duty cycle: Max. 5%, Max. 3min./hour
14