3. Use control push buttons to select and change the required parameter.
Right button [ 8 ]
Up Button [ h ]
Left button [ ESC ]
Down button [ i ]
4. Press [ENTER] to accept the parameter changes.
NOTE: Pressing [ESCAPE] returns unit to the upper menu.
5. Use control push buttons to:
Accept changes and leave menu
Do not accept changes and leave menu
Return to the System menu.
NOTE: Common relays are not configurable. Their condition is set to the following fixed settings:
Fixed Setting for Common Relays
• Alarm 1 on any channel de-energizes common Alarm 1 relay on the Main board.
• Alarm 2 on any channel de-energizes common Alarm 2 relay on the Main board.
• Fault on any channel de-energizes Fault relay on the Main board.
• System fault de-energizes Fault relay on the Main board.
5.3
General Setting Menu
The General Setting menu allows the user to:
• configure access passwords for the system and Calibration menu
• set parameters.