9.
Close lid securely. A properly closed lid will grasp the front silicon tab keeping the lid
closed. If a clicking sound is heard while the unit is operating, the lid is not fully closed.
Do not attempt to operate with the lid removed or open. Turn power switch ON. Unit will
not operate with lid open. If there is a smooth whirring sound and the unit accelerates
with little or no vibration, mySPIN 6 is ready to use. If there are loud or unusual sounds,
or excessive vibration, DO NOT OPERATE. Contact the Thermo Fisher Scientific Technical
Support.
10. To stop centrifugation, press power switch and wait until the rotor has completely stopped
rotating before opening the lid.
WARNING
Do not open lid until rotor has completely stopped! Avoid severe personal injury or property
damage from moving parts.
Rotor Replacement
To remove SnapSpin
center button firmly and at the same
time lift rotor
upwards.
To install SnapSpin
rotor onto the hex shaft.
Press center button firmly and
downward. Rotor should latch
with a click sound.
WARNING
Do not use damaged or cracked rotors!
8
SnapSpin Strip Tube
rotor: Press
rotor: Set