Maintenance; Regular Checks - Faro EDI Manual De Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

mAiNTENANCE

Replacing Front Screen (fig 4)
To remove Front Screen from Lamp proceed as follows:
• Turn screw on side of Screen counterclockwise using Tool supplied.
• Pull out Screen.
To reinstall Front Screen perform above operations in reverse.
Replacing bulb (fig 5/6)
(mod.halogen Lamp Gy 6.35 17 v 95 w)
To remove Bulb proceed as follows:
• Remove Front Screen (fig 4).
• Pull out Screen while pressing sides (fig 5).
• Withdraw Bulb "D" by pulling it upward (fig 6).
• Fit new Bulb. It must be as specified in the Data Sheet or on
Data Plate.
The new bulb should not be touched with the hands; handle it with
the protective shell. If it is accidentally touched with the fingers
wipe away the finger prints with cotton soaked in alcohol.
Replacing Fuse in Transformer model (fig 7).
The Lamp with Transformer has two Fuses of the same power. To replace them
proceed as follows.
• Make sure Power Cord is disconnected.
• Unscrew Plugs "E" located on Transformer case.
• Withdraw Fuses, check for burn out, replace if necessary.
new Fuses must have same rating as shown on data plate and in
Data Sheet.

REGULAR ChECkS

• Yearly
- Check mechanical soundness of joints and plastic elements.
- Legibility of label.
• Every two years
- Insulation test in accordance with EN 62353.
- Test for leakage in the casing.
- Check maximum and minimum illuminance.